progress
report 11/04/07
Martin Stute (martins@ldeo.columbia.edu)
Brice Loose (brice@ldeo.columbia.edu)
The objectives of this OTIC project are to develop a
sampler for dissolved gases that eliminates the need to collect and transport
water samples by extracting the gas from solution using membrane technology and
the principles of diffusion and solubility equilibrium. We have set three main criteria for the
sampler: (1) it should be able to
sample as many gases as possible (e.g. CFCs, SF6, SF5CF3,
3He, noble gases, N2, and CO2); (2) the
sampler should not require the use of a pure compressed carrier gas (N2 or He),
and (3) it should be portable, and have minimal power consumption. To date, we have made significant
progress on (1) and (2); equilibration experiments have proven that solubility
equilibrium can be achieved within the sampler for the gases N2, Ar,
SF6, O2 and CO2. Further work is necessary to achieve the same results with
CFCÕs and 3He. In this
report we will first outline and discuss the advances that have been completed
to date, including the limitations of the sampler, and then summarize by
listing the steps that are necessary to meet the three objectives described
above.

Figure 1. Schematic of the initial prototype developed in May, 2007.
To prove the design concept we have used
an experiment that takes advantage of the gases most present in the atmosphere,
N2 Ar and O2, while using CO2 as a more
variable gas. We equilibrate ca.
100 liters of water in a Nalgene tank with air by bubbling air through tank for
at least 8 hours. This reference
water volume is then pumped through the membrane sampler and a time series of
gas pressure and gas concentration is collected with syringes. To compare the gas concentration
yielded by the membrane sampler, syringe samples are collected from the water
volume and analyzed using headspace equilibration in the syringe. Both membrane and headspace samples
were analyzed on a Gas Chromatograph equipped with flame-ionization and thermal
conductivity detectors. We expect
the corrected concentration from the headspace samples to match the
concentration measured by the membrane sampler; as a further check, both of
these concentrations should reflect the concentration of the above gases in the
atmosphere (this is referred to as ÔDissolved AirÕ in Table 1). CO2
is more variable and subject to local sources, and reflects the ability of the
sampler to collect trace gas concentrations.
The initial prototype (Figure 1) consists of a Liqui-cel Mini-Module membrane whose
maximum water flow rate is 2.5 L/min.
A pressure transducer measures the gas pressure on the gas-side of the
membrane. Water pressure is
regulated via a flow valve, upstream of the membrane. The plumbing consists of nylon tubing with Parker o-ring
compression fittings. These
fittings are leak-tight and easy to assemble/disassemble.
RESULTS:
Table
1 shows the results of one of the equilibrium experiments. The ÔDissolved AirÕ concentrations are
those that we expect from equilibrium with the atmosphere. For each gas, the membrane sampler
yielded results within 3% of the expected values, and appears consistently more
accurate than the headspace method. The reproducibility of membrane samples is
very good, usually within 5%. Both
the headspace and membrane technique indicate CO2 concentrations
greater than in the atmosphere, which is not surprising as these experiments
were performed in a laboratory environment. The time to achieve solubility equilibrium is
well-captured by examining the time rate of change in the pressure
transducer. When equilibrium is
achieved, the pressure variations oscillated within several tenths of a mb of
the equilibrium value. The time to
equilibrium depends on the efficiency of the membrane, the dissolved gas
content of the water, and the flow rate.
At ca. 2 L/min, solubility equilibrium was achieved within 40-60
minutes. This implies that
approximately 80-120 liters of water are required to obtain a syringe sample of
~ 50 ml. Water that has greater
dissolved gas content will equilibrate more quickly; this was demonstrated this
summer during sampling at the LDEO borehole, where dissolved gas concentration
was much greater. We think that
the equilibration time is mostly a result of diffusive fractionation when the
rate of mass transfer is initially high (at low pressure). However, this should be less of a
concern for trace gases at low concentrations.
The same test was performed for CFC-11,
CFC-12 and CFC-113. We found that
CFC-11 was in good agreement, but the membrane produced values that were
50-100% greater than the headspace method. We expect this is an effect of the plastics, nylon and PTFE
in the system; these plastics are known to be a source of contamination to CFC
samples. We are currently
evaluating whether changing the plumbing to stainless steel can eliminate this
problem; the membrane housing and lumens are plastic, which may mean that CFCÕs
cannot be sampled in this fashion.
Table 1. Comparison between the headspace and membrane sampling
method, where both methods are expected to yield dissolved gas concentrations
that reflect solubility equilibrium of the water reservoir with the atmosphere
(referred to as Dissolved Air).
Except for CO2, both methods produced the expected results, within
8%. The membrane samples were
within 3% of expected values.
|
|
Ar (ml/L) |
N2 (ml/L) |
O2 (ml/L) |
CO2 (ml/L) |
|
Dissolved Air |
0.3242 |
12.3400 |
6.6150 |
0.3183 |
|
Headspace |
0.2649 |
12.2367 |
5.8368 |
0.5647 |
|
Membrane |
0.3238 |
12.6723 |
6.2409 |
0.6074 |
|
(Membrane/Air) |
0.99 |
1.027 |
0.943 |
1.908 |
Having proven the design concept, we are
refining the configuration and materials in an effort to include 3He as well as
CFCÕs in the list of gases which can be sampled with the membrane. As mentioned above, the membrane may
prove a source of CFCÕs. In the
case of 3He; we hope to minimize the effect of He diffusion through the walls
of the sampler by changing the nylon plumbing and fittings to stainless
steel. We intend to test the
sampler on water samples with dissolved gas content which are significantly
different from what is found in the atmosphere. This is necessary to confirm the integrity of the system and
demonstrate that the membrane will yield an accurate sample when dissolved gas
pressure is high.
The final phase of the design will be to
integrate the sampler into a portable suitcase-type configuration, to determine
the power consumption, and to evaluate second order effects, such as the
buildup of condensation on the gas side of the membrane. We anticipate that this work will be
completed by the early part of 2008, at which point we hope to deploy the
instrument in the long-term groundwater monitoring that is taking place in
Bangladesh. We will present a
poster on the results of this work at the 2007 AGU Fall Meeting in the General
Hydrology poster session.
We have a wiki that we have used to
aggregate our notes and to document progress on the sampler; http://duck-rabbit.ldeo.columbia.edu/PoTGas/wiki/index.php/Main_Page.