diff -r c39e37d59638 -r 5e67754f6457 UPLOOKER_pre_Firmware_16.3.cmd --- /dev/null Thu Jan 01 00:00:00 1970 +0000 +++ b/UPLOOKER_pre_Firmware_16.3.cmd Fri Jul 03 10:25:08 2020 -0400 @@ -0,0 +1,70 @@ +;====================================================================== +; U P L O O K E R _ P R E _ F I R M W A R E _ 1 6 . 3 . C M D +; doc: Tue Jun 15 11:46:07 2004 +; dlm: Fri Jul 3 10:08:36 2020 +; (c) 2004 A.M. Thurnherr +; uE-Info: 30 43 NIL 0 0 72 2 2 8 NIL ofnI +;====================================================================== + +; This is the uplooker command file for firmware pre 16.3 + +; NOTES: +; - contains only commands that change factory defaults +; - assumes that LADCP mode is installed +; - collect data in beam coordinates +; - single-ping ensembles; timing determined by [DOWNLOOKER.cmd] +; - narrow bandwidth +; - 25 8m cells --- reduce after determining the regional instrument +; range + +; HISTORY: +; Feb 1, 2006: +; - released for CLIVAR P16N +; Oct 30, 2006: +; - cleaned up for LADDER 1 +; Nov 19, 2008: +; - removed CR1 for acquire V1.3 +; Oct 14, 2009: +; - added explicit setting of narrow band mode +; Jul 3, 2020: +; - expunged master/slave terminology + +LP1 ; pings per ensemble +TP 00:00.00 ; time between pings +TE 00:00:00.00 ; time per ensemble + +LN25 ; number of depth cells +LS0800 ; bin size [cm] +LF0 ; blank after transmit [cm] + +WB1 ; narrow bandwidth mode 1 (not sure if required) +LW1 ; narrow bandwidth LADCP mode +LV250 ; ambiguity velocity [cm/s] + +SM2 ; receive sync pulses +SA011 ; wait for pulse before ensemble + +EZ0011101 ; Sensor source: + ; - manual speed of sound (EC) + ; - manual depth of transducer (ED = 0 [dm]) + ; - measured heading (EH) + ; - measured pitch (EP) + ; - measured roll (ER) + ; - manual salinity (ES = 35 [psu]) + ; - measured temperature (ET) + +EX00100 ; coordinate transformation: + ; - radial beam coordinates (2 bits) + ; - use pitch/roll (not used for beam coords?) + ; - no 3-beam solutions + ; - no bin mapping + +CF11101 ; Flow control: + ; - automatic ensemble cycling (next ens when ready) + ; - automatic ping cycling (ping when ready) + ; - binary data output + ; - disable serial output + ; - enable data recorder + +CK ; keep params as user defaults (across power failures) +CS ; start pinging