MASTER_pre_Firmware_16.3.cmd
changeset 12 5e67754f6457
parent 11 c39e37d59638
equal deleted inserted replaced
11:c39e37d59638 12:5e67754f6457
     1 ;======================================================================
       
     2 ;                    M A S T E R . C M D 
       
     3 ;                    doc: Tue Jun 15 11:46:07 2004
       
     4 ;                    dlm: Wed Oct 14 11:27:44 2009
       
     5 ;                    (c) 2004 A.M. Thurnherr
       
     6 ;                    uE-Info: 41 0 NIL 0 0 72 2 2 8 NIL ofnI
       
     7 ;======================================================================
       
     8 
       
     9 ; This is the default master/downlooker command file
       
    10 
       
    11 ; NOTES:
       
    12 ;   - contains only commands that change factory defaults
       
    13 ;   - assumes that LADCP mode is installed
       
    14 ;   - collect data in beam coordinates
       
    15 ;   - staggered single-ping ensembles every 1.5s/2.0s
       
    16 ;   - narrow bandwidth
       
    17 ;   - 25 8m cells --- reduce after determining the regional instrument
       
    18 
       
    19 ; HISTORY:
       
    20 ; Feb  1, 2006:
       
    21 ;	- released for CLIVAR P16N
       
    22 ; Oct 30, 2006:
       
    23 ;	- adapted for LADDER 1
       
    24 ; Nov 19, 2008:
       
    25 ;	- removed CR1 for acquire V1.3
       
    26 ; Oct 14, 2009:
       
    27 ;	- added explicit setting of narrow band mode
       
    28 
       
    29 TC2			; ensembles per burst
       
    30 LP1			; pings per ensemble
       
    31 TB 00:00:03.50		; time per burst
       
    32 TE 00:00:01.50		; time per ensemble
       
    33 TP 00:00.00		; time between pings
       
    34 
       
    35 LN25			; number of depth cells
       
    36 LS0800			; bin size [cm]
       
    37 LF0			; blank after transmit [cm]
       
    38 
       
    39 WB1			; narrow bandwidth mode 1 (not sure if required)
       
    40 LW1			; narrow bandwidth LADCP mode
       
    41 LV250			; ambiguity velocity [cm/s]
       
    42 
       
    43 SM1			; master
       
    44 SA011			; send pulse before each ensemble
       
    45 SW5500			; wait .5500 s after sending sync pulse
       
    46 SI0			; # of ensembles to wait before sending sync pulse
       
    47 
       
    48 EZ0011101		; Sensor source:
       
    49 			;	- manual speed of sound (EC)
       
    50 			;	- manual depth of transducer (ED = 0 [dm])
       
    51 			;	- measured heading (EH)
       
    52 			;	- measured pitch (EP)
       
    53 			;	- measured roll (ER)
       
    54 			;	- manual salinity (ES = 35 [psu])
       
    55 			;	- measured temperature (ET)
       
    56 
       
    57 EX00100			; coordinate transformation:
       
    58 			;	- radial beam coordinates (2 bits)
       
    59 			;	- use pitch/roll (not used for beam coords?)
       
    60 			;	- no 3-beam solutions
       
    61 			;	- no bin mapping
       
    62 
       
    63 CF11101			; Flow control:
       
    64 			;	- automatic ensemble cycling (next ens when ready)
       
    65 			;	- automatic ping cycling (ping when ready)
       
    66 			;	- binary data output
       
    67 			;	- disable serial output
       
    68 			;	- enable data recorder
       
    69 
       
    70 CK			; keep params as user defaults (across power failures)
       
    71 CS			; start pinging