-;                    M A S T E R . C M D 
-;                    doc: Tue Jun 15 11:46:07 2004
-;                    dlm: Wed Oct 14 11:27:44 2009
-;                    (c) 2004 A.M. Thurnherr
-;                    uE-Info: 41 0 NIL 0 0 72 2 2 8 NIL ofnI
-;======================================================================
-
-; This is the default master/downlooker command file
-
-; NOTES:
-;   - contains only commands that change factory defaults
-;   - assumes that LADCP mode is installed
-;   - collect data in beam coordinates
-;   - staggered single-ping ensembles every 1.5s/2.0s
-;   - narrow bandwidth
-;   - 25 8m cells --- reduce after determining the regional instrument
-
-; HISTORY:
-; Feb  1, 2006:
-;	- released for CLIVAR P16N
-; Oct 30, 2006:
-;	- adapted for LADDER 1
-; Nov 19, 2008:
-;	- removed CR1 for acquire V1.3
-; Oct 14, 2009:
-;	- added explicit setting of narrow band mode
-
-TC2			; ensembles per burst
-LP1			; pings per ensemble
-TB 00:00:03.50		; time per burst
-TE 00:00:01.50		; time per ensemble
-TP 00:00.00		; time between pings
-
-LN25			; number of depth cells
-LS0800			; bin size [cm]
-LF0			; blank after transmit [cm]
-
-WB1			; narrow bandwidth mode 1 (not sure if required)
-LW1			; narrow bandwidth LADCP mode
-LV250			; ambiguity velocity [cm/s]
-
-SM1			; master
-SA011			; send pulse before each ensemble
-SW5500			; wait .5500 s after sending sync pulse
-SI0			; # of ensembles to wait before sending sync pulse
-
-EZ0011101		; Sensor source:
-			;	- manual speed of sound (EC)
-			;	- manual depth of transducer (ED = 0 [dm])
-			;	- measured heading (EH)
-			;	- measured pitch (EP)
-			;	- measured roll (ER)
-			;	- manual salinity (ES = 35 [psu])
-			;	- measured temperature (ET)
-
-EX00100			; coordinate transformation:
-			;	- radial beam coordinates (2 bits)
-			;	- use pitch/roll (not used for beam coords?)
-			;	- no 3-beam solutions
-			;	- no bin mapping
-
-CF11101			; Flow control:
-			;	- automatic ensemble cycling (next ens when ready)
-			;	- automatic ping cycling (ping when ready)
-			;	- binary data output
-			;	- disable serial output
-			;	- enable data recorder
-
-CK			; keep params as user defaults (across power failures)
-CS			; start pinging